Home Syringe Pumps/Peristaltic Pumps Constant Pressure Syringe Pumps Remote PHD ULTRA™ CP 4400 Syringe Pump


Larger Views

 

Remote PHD ULTRA™ CP 4400 Syringe Pump

The PHD ULTRA™ CP Syringe Pump is the first of its kind to allow operation under constant pressure or constant flow. Historically, the only means available to dispense at constant pressure has involved the use of various amplifiers and other accessories/software. The PHD ULTRA™ CP Syringe Pump, when combined with virtually any commercially available pressure transducer/amplifier combination with 0-10V DC analog output, results in a constant pressure dispensing system. This sytem can deliver fluids with an applied force up to 1,000 lbs (depending upon the pump).

Using Harvard Apparatus syringe pump technology and software-controlled pressure monitoring, the PHD ULTRA™ CP Syringe Pump is able to maintain a user-defined system pressure ±2%. A user-adjustable sensitivity setting allows for the customization of the system response time necessary to attain the set pressure. The PHD ULTRA™ CP Syringe Pump allows the user to set pressures in the units of their choice including mmHg, kPa, and psi.

Search for Product in current catalog
Item Listing Full Description Specifications Journal Articles Manuals
Item# Description
883316 PHD ULTRA™ CP Remote 4400 Syringe Pump Infuse/Withdraw Programmable

For local pricing, contact your nearest distributor.



Full Description Back to Top

Features:
• Enables continuous pressure-controlled infusion
• Use in constant flow or constant pressure mode
• Enhances safety for sensitive infusion targets and physiological experimentation
• Compatible with a wide range of pressure transducers
• Automatically adjusts flow rate to maintain constant pressure
• Alpha/numeric keyboard without a PC
• Icon operation
• New color LCD touch screen
• Up-front control knobs for ease of operation
• Vertical or horizontal orientation
• Adjustable linear force to 200 lbs across the entire flow range
• CE, ETL (UL, CSA), WEEE, EU RoHS & CB Scheme Approved
• 2 year warranty

Applications:
• Short-term organ perfusion studies that require the maintenance of physiologic conditions.
• Introduction of chemical reactants in a controlled manner.
• Ocular injections and perfusions.
• Small animal whole-body perfusions.
• Constant pressure microfluidic mixing.
• Administration of genetic material into organs without viral vectors.

The PHD ULTRA™ CP Syringe Pump is the first of its kind to allow operation under constant pressure or constant flow. Historically, the only means available to dispense at constant pressure has involved the use of various amplifiers and other accessories/software. The PHD ULTRA™ CP, when combined with virtually any commercially available pressure transducer/amplifier combination with 0-10V DC analog output, results in a constant pressure dispensing system. This sytem can deliver fluids with an applied force up to 1,000 lbs (depending upon the pump).

Superior Functionality
Using Harvard Apparatus syringe pump technology and software-controlled pressure monitoring, the PHD ULTRA™ CP Syringe Pump is able to maintain a user-defined system pressure ±2%.

A user-adjustable sensitivity setting allows for the customization of the system response time necessary to attain the set pressure. The PHD ULTRA™ CP Syringe Pump allows the user to set pressures in the units of their choice including mmHg, kPa, and psi.

The PHD ULTRA™ CP Remote Syringe Pump consists of a control unit and syringe pumping mechanism all connected via an RS-485 (IEEE-1394) cable. This pump includes a 30 foot RS-485 (IEEE-1394) cable allowing the syringe pumping mechanism and the control box to be separated. Use the remote pump in hazardous environments where the researcher is safer when distanced from the material being pumped or for applications where it is necessary to have your pump inside an incubator.

Essential Application Tools
The PHD ULTRA™ CP Syringe Pump can work with any pressure transducer and amplifier that will output a 0-10V or similar signal. We offer a wide variety of pressure transducers and amplifiers to suit your particular application. Our most popular physiological pressure transducer is our APT-300, a fluid-filled unit with a pressure range of ±300mmHg and a sensitivity of 5µV/V/mmHg (±1%). This has an easily replaceable transducer head. For low pressure applications, such as maintaining the pressure in an eye, we have the P75 Venous Pressure Transducer. This pressure transducer has a pressure range of ˝75mmHg and a sensitivity of 1mV/mmHg. When the customer requires a separate amplifier, the unit of choice is the model 601 Transducer AmplifierModule, or TAM-A. As a PLUGSYS module, this DC Bridge Amplifier requires the use of a PLUGSYS housing with a power supply. Cases holding 2, 5, or 10 modules are available.

Pressure & Flow Rate Data
Data can be monitored via RS-232 from the PHD ULTRA™ CP Syringe Pump to a PC. In practice, the 0-10V analog output of a pressure transducer amplifier is connected to the analog input on the rear panel of the pump. The amplifier or signal conditioner can be provided by the customer, or accomplished with various tranducers and amplifiers available from Harvard Apparatus. The pressure range may be scaled to fit the available transducer voltage output for systems that output less than 10 V. While in constant pressure mode, in addition to the set and actual pressure, the pump displays the flow rate. This data may be output for further analysis with a variety of data acquisition packages.

Accuracy & Reproducibility
In addition to constant pressure mode, the PHD ULTRA™ CP may also be used in flow mode with its worldrenowned accuracy and reproducibility.
All PHD ULTRA™ CP Syringe Pumps are infuse/withdraw programmable models. When used in flow mode, these offer programmable features such as method storage and flow programming functions to allow the user to create simple to complex methods. The PHD ULTRA™ CP Syringe Pump is available in a variety of configurations to suit the desired pressure or flow rate ranges.

Advanced Programming Features:
• Flow Programming - Change the flow with time, volume or a triggered event as many times as you like
• Bolus - Inject a drug (or drugs) in a high quantity at once. The bolus injection can be made in time or volume.
• Concentration Delivery - Calibrate flow in concentration units of mg/kg easily so flow is calibrated to concentration of drug and animal weight.
• Gradients
• % ratio - up to three solvents
• I/O — dedicated and user defined I/O
• Pulsed Flow- so you can program the pulse easily

Widest Flow Rate Range
This pump is engineered to provide flow accuracy within 0.35% and reproducibility within 0.05%. One syringe from 0.5 µl to 140 ml pump at a range of 3.06 pl/min to 215.8 ml/min.

Maximum Experimental Versatility
The pump can be oriented vertically or horizontally for optimum experimental connectivity. This pump comes standard to hold 1 syringe.

Easy-to-Use Interface
The PHD ULTRA™ color LCD touch screen graphic interface is divided into three basic areas: Operations Display, Message Area, and Navigation. This configuration allows you to easily move through all menu selections and data entry by gently touching the onscreen buttons with a finger or the tip of a soft, non-sharp object such as a pencil eraser.

The Methods Main or Quick Start screens are the primary ?home? for the applications. From those screens you access all the commands needed to operate the PHD ULTRA™ as well as the main system settings.

The Message Area of the touch screen is used to display helpful instructions for the currently displayed screen. It is also used to display error or warning messages to indicate problem conditions in a Method or error conditions during pump operation.

The software is organized into three main Navigational branches, the quick start operations, preloaded/user-defined Methods, and systems settings. You can control operations directly with the touch screen or remotely from an independent computer or device via the external I/O interface.

Program Description
To operate the PHD ULTRA™ the user defines all the required parameters for infusing or withdrawing liquids through a Pump Control Method. The basic operation is a simple 4-step procedure:

1. Select a method
2. Enter operating parameters
3. Preview your method
4. Run your method

Applications
• Animal Infusions or Withdrawals - The PHD ULTRA™ will control the delivery of varying % of nutrients or drugs infused into animals, flush lines using catheters, needles, cannulae or microdialysis.
• Proportioning and Delivering of Mixtures - Mixing gradients or proportions with independent control of two liquids.
• Aerosol for Coating - The pump at high pressure can create an aerosol for the delivery of coating materials such as pharmaceutical tablets and aerosol studies.
• Delivery to Mass Spectroscopy - The delivery of fluids to the MS for calibration, matrix addition or ESI sample.
• Compensating Flows - The continuous infusion and simultaneous withdrawal of liquids for cell cultures or perfusion chambers.
• Dispensers/Injectors — Adhesives, Cell injection, MRI Dyes, Activators/Enzymes, Flow injection, Microreaction vessels, or Stereotaxic delivery.

Advanced GLP Documentation Features:
• Experiment parameter download information to PC
• Alpha/numeric capability

Accessories
A full range of accessories are compatible with the PHD ULTRA™ including syringe heaters, in-line heaters and coolers, nanofluidic circuits, connectors, tubing, syringes and more.


Specifications Back to Top

Specifications883316
Pump ConfigurationRemote
Pump FunctionInfuse/Withdraw Programmable
Flow Rate Maximum215.8 ml/min using 140 ml syringe
Flow Rate Minimum3.06 pl/min using 0.5 µl syringe
No of Syringes1
Syringe Rack TypeStandard Rack
Syringe Size Minimum0.5 µl
Syringe Size Maximum140 ml
Max Linear Force200 lbs @ 100% Force Selection
I O TTL Connectors15 pin D-Sub Connector
RS 232 Connectors9 pin D-Sub Connector
USB ConnectorsType B
Flow Accuracy±0.35%
Pressure Accuracy±2%
Reproducibility±0.05%
Non Volatile MemoryStorage of all settings
Drive Motor1.8° Stepper Motor
Motor Drive ControlMicroprocessor with 1/16 microstepping
Number of Microsteps per one rev of Lead Screw6,400
Step Rate Minimum27.5 sec/µstep
Step Rate Maximum52 µsec/µstep
Pusher Travel Rate Minimum0.36 µm/min
Pusher Travel Rate Maximum190.8 mm/min
Display4.3" WQVGA TFT Color Display with Touchpad
Input Power75 W, 0.5 A fuse
Voltage Range100-240 VAC, 50/60 Hz
Dimensions Control Box LxDxH in cm 12.0 x 8.5 x 7.25 (30.48 x 21.59 x 18.4)
Dimensions Remote Box LxDxH in cm 12.0 x 5.5 x 7. (30.5 x 14.0 x 17.8)
Net Weight Metric6.4 kg
Net Weight English14.1 lb
Environmental Operating Temperature Metric4°C to 40°C
Environmental Operating Temperature English40°F to 104°F
Environmental Storage Temperatue Metric-10°C to 70°C
Environmental Storage Temperatue English14°F to 158°F
Environmental Humidity20% to 80% RH, non condensing
Mode of OperationContinuous
ClassificationClass I
Pollution Degree1
Installation CategoryII
Regulatory CertificationsCE, ETL, (UL, CSA), WEEE, EU RoHS & CB Scheme

Back to Top



Journal Articles Back to Top

Displaying 1 to 10 of 102results for Remote PHD ULTRA™ CP 4400 Syringe Pump   Page 1  2  3  4  5  6  7  8  9  10  11 

Score

Document

Description

100% Login to view PP117 phd 22 2000.pdf
[1.4 MB]
PHD 22_2000_forpdf.qxd
94% Login to view PP116 phd 2.pdf
[90 KB]
PHD 2000 Programmable Syringe Pump used in Analysis of Biomolecular Interactions Using a Miniaturized Surface Plasmon Resonance Sensor
75% Login to view PP96 Intrarenal Dopamine Production and Distribution in the Rat.pdf
[399 KB]
Calibrated Syringe Pump use in finding Anticonvulsant Properties of Ganazolone
75% Login to view PP114 pump microchannel.pdf
[74 KB]
Syringe Infusion Pump use in finding Microchannel Devices for Efficient Contacting of Liquids in Solvent Extraction
73% Login to view PP69 PHD 2000 Syringe Pump.pdf
[598 KB]
Passive mixing in a three-dimensional serpentine microchannel - Microelectromechanical Systems, Journal of
72% Login to view PP06 PHD2000 Syringe Pump.pdf
[104 KB]
PHD 200 Syringe Pump use in finding Molecular Clocks
46% Login to view PP53 Syringe Pump22.pdf
[525 KB]
Syringe Pump use in finding Choroid plexus epithelial expression of MDRI P glycoprotein
41% Login to view PP36 Syringe Pump.pdf
[133 KB]
Surface Micromachine Microfluidics: Design, Fabrication, Packa
41% Login to view PP06 Molecular Clocks.pdf
[108 KB]
PHD 22/2000 Syringe Pump used in finding Molecular Clocks
37% Login to view PP04 pump ms.pdf
[53 KB]
Analysis of Jojoba Oil with syringe pump 'apparatus 22'
   

Back to Top



Manuals Back to Top

Displaying 1 to 3 of 3results for Remote PHD ULTRA™ CP 4400 Syringe Pump  

Score

Document

Description

100% Login to view 703005 PHD ULTRA Manual.pdf
[2.4 MB]
PHD Ultra Syringe Pumps Manual - Standard/Remote/Programmable
100% Login to view 703005 PHD ULTRA Quickstart.pdf
[173 KB]
PHD Ultra Syringe Pumps Quickstart Guide - Standard/Remote/Programmable
54% Login to view PHD2000.pdf
[493 KB]
PHD 2000 Syringe Pump Series Manual
   

Back to Top

Breakthrough Products for Regenerative Medicine
2011 Regenerative Medicine Brochure
100's of Innovative products developed with leading scientists & surgeons!
  • Injection & Perfusion
  • Bioreactors
  • Biosensing
  • Total Solutions
  • Total Support


Announcing the new
PHD ULTRA™ Syringe Pumps

Click here for more information about PHD ULTRA™ Syringe Pumps and to download a brochure.

  • Patent pending mechanical design provides highest accuracy/precision & smoothest flow
  • Easy operation with advanced touch screen
  • Advanced EZ PRO™ software for PC functions without a PC.


Sign-up for a FREE Full Line Bioresearch Catalog
CLICK HERE!

  • Over 11,000 specialty products to enhance your bioresearch
  • Enhanced solutions for small animal research
  • 100's of pages of specialized information on instrumentation, applications, uses of products

 

Products on this website are for Research Use Only. Not for use in humans unless
proper investigational device regulations have been followed.


Privacy Policy   |   Terms and Conditions   
Copyright© Harvard Apparatus Ltd. - All rights reserved.
(44) 1732-864001
Website-related comments or concerns can be emailed to our webmaster.

Anesthesia Equipment Animal Handling Products Behavior Research Products Biological Measurement Products Cardiovascular Products Cell Biology Products Microincubation / Microinjectors CPK« Atomic Models Delivery & Sampling Products Dialysis / Sample Prep / SPE Electroporation / Transfection Electrophysiology Products General Laboratory Products Homeothermic/Temperature Control Isolated Organ & Isolated Tissue Molecular Biology Products Neuroscience Products Non-Invasive Blood Pressure Physiology Products Respiratory Products Stereotaxic Instruments Surgical Equipment Surgical Instruments Syringe Pumps/Peristaltic Pumps Tubing & Connectors Ventilators