Disposable Blood Pressure Transducer for PowerLab
The MLT0670 is a low-cost, light-weight, disposable blood pressure transducer that gives consistent and accurate readings. The transducer is supplied in sterile packaging, however MUST be sterilized prior to use with human subjects. The transducer is suitable for use with the FE117 BP Amp using the MLAC05 Adapter Cable.
The MLT0670 is a disposable blood pressure transducer designed for direct measurement of arterial and venous pressure (-50 to +300 mmHg). It is a low-cost, light-weight, disposable blood pressure transducer that gives consistent and accurate readings. The transducer is suitable for use with the FE117 BP Amp using the MLAC05 Adapter Cable.
Operation
Before use, finger-tighten all connections. Blood pressure is transmitted to the sensor via a subject cannula or catheter. Before connecting to a catheter, prime the system by flushing three or four times until all visible air bubbles are removed. Use a 10 mL syringe if continuous flushing is not available. Open the 3-way stop cock to air to take a zero reading for the transducer before calibration and recording blood pressure.
Application
The MLT0670 Disposable BP Transducer is designed for measuring arterial and venous blood pressure in humans and large animals.The transducer is supplied in sterile packaging, however MUST be sterilized prior to use with human subjects.
The MLT0670 is a disposable blood pressure transducer designed for direct measurement of arterial and venous pressure (-50 to +300 mmHg) in humans (the transducer is supplied in sterile packaging, however MUST be cold-sterilized before use with humans). The transducer can also be used in animal applications. The MLT0670 is ONLY for single use in humans and must be used with the FE117 BP Amp and MLAC05 Adapter Cable (not included).
Specifications | 77-0064 |
---|---|
Cable Length English | 9.8 in |
Cable Length Metric | 250 mm |
Certifications | *na |
Connector | Spacelabs plug |
Control Port | *na |
Dimensions H x W x D English | *na |
Dimensions H x W x D Metric | *na |
Excitation | 2 to 10 V DC |
Frequency Response | *na |
Input Accuracy | *na |
Input Amplification Ranges | *na |
Input Connection Type | *na |
Input Impedance | 350 Ω ± 10% |
Input Impedance | See product page for complete specifications |
Input Isolation Rating | *na |
Input Isolation | *na |
Input Leakage Current | *na |
Input Safety | *na |
Input Sensitivity | *na |
Input User Offset | *na |
Input Zeroing and Offset | *na |
Length English | 3.4 in |
Length Metric | 85 mm |
Maximum Power Needs | *na |
Model | MLT0670 |
Non linearity and Hysteresis | ±2 % of reading or ±1 mmHg (whichever is greater) |
Operating Conditions | *na |
Operating Life | >500 hours |
Operating Pressure Range | ?50 to +300 mmHg |
Operating Temperature | 15 °C to +40 °C |
Output Analog Signal | *na |
Output Impedance | <400 Ω |
Overpressure protection | -400 to +4000 mmHg |
Package | 1 |
Product Family | Data Acquisition |
Sensitivity | 5 ?V/V/mmHg |
Storage Temperature | -25 °C to +70 °C |
Transducer Alternative Types | *na |
Transducer Excitation | *na |
Transducer Type | *na |
Weight English | *na |
Weight Metric | *na |
Weight | (including cable): 26 g |
Width English | 1.6 in |